The authors developed a high-frequency chest compression (HFCC) device to aid in mucous clearance for patients with obstructive lung disease. The device, designed for self-therapy, consists of a large-volume variable-frequency air-pulse delivery system and a nonstretchable inflatable vest worn by the patient. Pressure pulses are controlled by the patient and applied during expiration. Pulse frequency is tunable from 5 to 25 Hz. Maximum vest pressure is 39 mmHg (5.2 kPa), with patient-controlled vest inflation and deflation time constants of 0.5 s. Vest pressure increases from 28 mmHg (3.7 kPa) at 5 Hz to 39 mmHg (5.2 kPa) at 25 Hz. Preliminary clinical trials have shown the HFCC device to be more effective than standard chest physical therapy. The HFCC device yielded a mean volume of cleared mucus of 3.3 cc per session, compared with 1.8 cc for a conventional therapy session.