Table 2

 Comparative results between the Suzanne and Remmers monitors

SuzanneRemmersp Value*
*Wilcoxon signed rank test.
Values are median (interquartile range).
RDI, respiratory disturbance index measured as the number of events/h recording time; Spo2 mean, mean oxygen saturation measured by pulsed oximetry over the recording time period; Spo2 min, minimum oxygen saturation recorded by pulse oximetry during the recording time period; ΔSpo2, arithmetic difference between Spo2 mean and Spo2 min.
RDI (n/h)12.9 (5.4–28)8.8 (2.9–18.2)<0.001
Recording time (min)431 (379–496)429 (381–486)<0.022
Spo2 mean (%)95.9 (94.9–96.8)95.1 (93.9–96.1)<0.001
Spo2 min (%)88.5 (84–91)85 (80–87)<0.001
ΔSpo2 (%)7.7 (4.8–11.1)10.4 (7.9–13.9)<0.001
% time Spo2 <90% (%)0.15 (0–1.4)0.6 (0.1–1.6)<0.001