RT Journal Article SR Electronic T1 Measurement of twitch transdiaphragmatic pressure: surface versus needle electrode stimulation. JF Thorax JO Thorax FD BMJ Publishing Group Ltd and British Thoracic Society SP 669 OP 670 DO 10.1136/thx.46.9.669 VO 46 IS 9 A1 A Mier A1 C Brophy YR 1991 UL http://thorax.bmj.com/content/46/9/669.abstract AB The transdiaphragmatic pressure (Pdi) generated during bilateral supramaximal phrenic nerve stimulation at 1 Hz from surface stimulating electrodes was compared with pressures obtained from needle electrodes inserted under local anaesthesia. Surface electrodes were used to obtain diaphragmatic electromyograms and magnetometers to monitor rib cage and abdominal configuration. Twitch Pdi was recorded at functional residual capacity in three normal subjects. Mean (SD) twitch Pdi in the three subjects during stimulation with surface electrodes was 19.4 (1.8), 22.5 (1.1), and 29.3 (2.2) cm H2O compared with 12.9 (1.5), 17:4 (1.3), and 22.6 (3.0) cm H2O with needle stimulating electrodes. Thus phrenic nerve stimulation with needle electrodes was more complicated and more invasive than stimulation with surface electrodes and resulted in lower transdiaphragmatic pressures.