Exhaled nitric oxide measurements: clinical application and interpretation
Free
  1. D R Taylor1,
  2. M W Pijnenburg2,
  3. A D Smith1,
  4. J C D Jongste2
  1. 1Department of Medical and Surgical Sciences, Dunedin School of Medicine, University of Otago, P O Box 913, Dunedin, New Zealand
  2. 2Department of Paediatric Respiratory Medicine, Erasmus MC - Sophia Children’s Hospital, University Medical Centre, P O Box 2060, 3000 CB Rotterdam, The Netherlands
  1. Correspondence to:
    Professor D R Taylor
    Department of Medical and Surgical Sciences, Dunedin School of Medicine, University of Otago, P O Box 913, Dunedin, New Zealand; robin.taylor{at}stonebow.otago.ac.nz

Abstract

The use of exhaled nitric oxide measurements (FEno) in clinical practice is now coming of age. There are a number of theoretical and practical factors which have brought this about. Firstly, FEno is a good surrogate marker for eosinophilic airway inflammation. High FEno levels may be used to distinguish eosinophilic from non-eosinophilic pathologies. This information complements conventional pulmonary function testing in the assessment of patients with non-specific respiratory symptoms. Secondly, eosinophilic airway inflammation is steroid responsive. There are now sufficient data to justify the claim that FEno measurements may be used successfully to identify and monitor steroid response as well as steroid requirements in the diagnosis and management of airways disease. FEno measurements are also helpful in identifying patients who do/do not require ongoing treatment with inhaled steroids. Thirdly, portable nitric oxide analysers are now available, making routine testing a practical possibility. However, a number of issues still need to be resolved, including the diagnostic role of FEno in preschool children and the use of reference values versus individual FEno profiles in managing patients with difficult or severe asthma.

  • AHR, airway hyperresponsiveness
  • CF, cystic fibrosis
  • COPD, chronic obstructive pulmonary disease
  • Feno, exhaled nitric oxide
  • FEV1, forced expiratory volume in 1 second
  • ICS, inhaled corticosteroids
  • NO, nitric oxide
  • PCD, primary ciliary dyskinesia
  • exhaled nitric oxide
  • asthma
  • diagnosis
  • monitoring
  • treatment

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

It is now 12 years since it was first reported that exhaled nitric oxide (NO) levels are increased in bronchial asthma.1 This discovery followed a period of intense interest in the biology of NO during the late 1980s.2 The numerous roles of NO in respiratory pathophysiology have been extensively reviewed.3 NO is an endogenous messenger with a diverse range of effects including non-adrenergic, non-cholinergic neurotransmission, vascular and non-vascular smooth muscle relaxation.3

There is contradictory evidence regarding the exact function of NO in lung disease. In pathological situations NO is a pro-inflammatory mediator with immunomodulatory effects.3 This appears to predispose to the development of airway hyperresponsiveness (AHR), although this is not a consistent finding.3,4 On the other hand, under physiological conditions NO acts as a weak mediator of smooth muscle relaxation and protects against AHR.5 In exhaled air NO appears to originate in the airway epithelium.6 Although raised levels may occur with a number of airway or lung diseases,7 the most important context in which the measurement of NO is clinically useful is that of allergic airways disease.

It is against this background that measuring the fraction of NO in exhaled air (Feno) has emerged as a potentially important clinical tool. Feno can be measured easily using a range of commercially available analysers, and smaller less costly devices are now becoming available. This opens the possibility that Feno measurements might be used routinely in the assessment of airway disease. This is a significant advance. To date, assessing airway physiology—that is, changes in airway calibre and/or bronchodilator or bronchoconstrictor responsiveness—has been the principal means of providing supportive evidence for the diagnosis of airways disease and assessing severity. Although pulmonary function tests will always remain important, they are one step removed from the issue of interest—that is, airway inflammation. Thus, Feno measurements provide a complementary and, in some instances, a more relevant perspective.

In this paper we will address key issues of importance to both adult and paediatric respiratory clinicians who are contemplating using Feno measurements in day to day practice.

RATIONALE FOR THE USE AND INTERPRETATION OF FEno MEASUREMENTS

A number of lines of evidence converge to provide the rationale for using Feno measurements in the assessment and management of respiratory disease. There are two key points: (1) there is a highly significant relationship between Feno and eosinophilic airway inflammation, and (2) there is an equally important relationship between eosinophilic airway inflammation and steroid responsiveness. The evidence is summarised as follows:

  • Feno measurements are highly correlated with eosinophilic airway inflammation.

  • Eosinophilic airway inflammation is associated with a positive response to steroid treatment.

  • Raised Feno levels predict steroid responsiveness in patients with non-specific respiratory symptoms.

  • The use of inhaled corticosteroid (ICS) treatment in asthma results in a fall in Feno, and there is a dose-dependent relationship between ICS and Feno.

Feno measurements are highly correlated with eosinophilic airway inflammation

Atopic asthma is characterised by an inflammatory infiltrate in the airways, with a predominance of mast cells and eosinophils.8 Studies confirm that Feno measurements correlate well with airway eosinophilia in induced sputum,9,10 biopsy material,11–13 and bronchoalveolar lavage fluid.14 In one study a significant relationship between Feno and blood eosinophils was also reported.15 A similar relationship has been described between Feno and sputum eosinophilic cationic protein16 in patients with asthma.

Importantly, two studies have shown that the relationship between Feno levels and airway eosinophilia is independent of the clinical diagnosis. It has been reported in patients with chronic obstructive pulmonary disease (COPD).17 In the study by Brightling et al11 patients who did not fulfil criteria for the diagnosis of asthma but who had eosinophilic bronchitis had raised Feno levels. In atopic patients with allergic rhinitis but no asthma, Feno levels are also raised.18–20 Similarly, in atopic asthmatic subjects in remission for many years, but who nevertheless have eosinophilic airway inflammation in bronchial biopsies, Feno levels are increased.13 All of these data form the basis on which Feno measurements are considered reliable as a non-invasive marker of eosinophilic airway inflammation.

Eosinophilic airway inflammation is associated with a positive response to steroid treatment

Treatment with corticosteroids results in a reduction in airway eosinophilia in asthma and a simultaneous improvement in clinical parameters.21,22 In contrast, in asthma which is not characterised by eosinophilia (at least in sputum), the response to steroids is likely to be poor.23,24 These findings also apply in patients with fixed airflow obstruction in whom neither the history nor physiological measurements permit easy discrimination between asthma and COPD. In such patients, a positive outcome with a trial of steroid treatment is associated with the presence of sputum eosinophilia.25,26 Thus, assessing the character of airway inflammation (eosinophilia) appears to be important in the initial management of patients with chronic respiratory symptoms in order to identify those who are more likely to benefit from treatment with steroids.

Raised Feno levels predict steroid responsiveness in patients with non-specific respiratory symptoms

Little et al27 have shown that the clinical benefit of increased steroid treatment in patients with asthma is greatest in patients with raised Feno levels. This has been taken a step further by Smith et al28 who evaluated the predictive accuracy of Feno measurements (as a surrogate for airway eosinophilia) in adult patients with undiagnosed respiratory symptoms. In that study, the positive and negative predictive values for a range of outcomes following a trial of inhaled fluticasone were superior for Feno as a predictor than spirometry, bronchodilator response, and measurements of AHR. Importantly, this study identified an optimum cut point for steroid response at an Feno of 47 ppb (fig 1). This outcome was largely independent of the final diagnosis. A similar result has been reported by Szefler et al29 who showed that children with high Feno values are more likely to respond to ICS than children with lower Feno values.

Figure 1

 Steroid responsiveness in relation to Feno measurements in patients with non-specific chronic respiratory symptoms. Mean (SE) changes from baseline in (A) forced expiratory volume in 1 second (FEV1), (B) morning peak flow over last 7 days of treatment, (C) composite symptom score, and (D) provocative concentration of adenosine monophosphate causing a 20% fall in FEV1 (PC20AMP) following treatment with inhaled fluticasone 500 µg/day (minus change with placebo), stratified by baseline Feno expressed as tertiles. Comparisons between tertiles were performed using one way analysis of variance with linear contrasts to identify any trend across the three tertiles; *p<0.05; †p<0.01; ‡p<0.001. Reproduced from Smith et al28 with permission of the publishers.

ICS treatment in asthma results in a fall in Feno with a dose-dependent relationship between ICS and Feno

A number of studies have shown that ICS treatment results in a fall in Feno levels in patients with mild asthma.30–32 Both the magnitude and the time interval over which the reduction occurs are dose-dependent33,34 and the response is reproducible.35 Feno levels tend to plateau at higher doses of ICS.36 In addition, there are highly significant correlations between the changes in Feno and changes in induced sputum eosinophils with ICS therapy.37

Taken together, these data provide foundational evidence that Feno measurements have a potentially important role in evaluating and treating patients with airways disease. Firstly, Feno may be used as a surrogate marker for airway diseases characterised by eosinophilia such as atopic asthma, cough variant asthma, and eosinophilic bronchitis. Secondly, because of the close relationship between steroid responsiveness and airway eosinophilia (in contrast to other histological phenotypes), Feno measurements have a role in predicting and monitoring the response to ICS treatment.

DIAGNOSING AIRWAYS DISEASE

Establishing a diagnosis is the first step in clinical management but, for diseases of the airways, a diagnostic label has its limitations. The term “chronic obstructive airways disease” (COPD) encompasses a spectrum of overlapping pathologies and the phenotype is a mixed one, especially in relation to treatment. The same is true for bronchial asthma, which is increasingly acknowledged to be heterogeneous,38,39 particularly if it is severe.40,41 Against this background, and given the specificity of Feno measurements as a marker for eosinophilic airway inflammation, it is not surprising that Feno offers advantages as well as limitations as a “test for asthma”.

Asthma

In adults, Feno measurements are helpful in discriminating asthma from non-asthma.42 It is best to reserve the test for investigating chronic symptoms (of 6 weeks duration or longer) because viral illness may give rise to a false positive result.43,44 In the study by Dupont et al,45 among 240 non-smoking steroid naïve individuals of whom 160 (67%) fulfilled the criteria for the diagnosis of asthma, Feno levels were highly predictive of asthma with a sensitivity and specificity of 85% and 90%, respectively. In the study by Smith et al,46 similar sensitivity (88%) and specificity (79%) were obtained in 47 patients of whom 17 had asthma. Predictive values were almost identical to those obtained using induced sputum cell counts. A striking feature in that study was the poor performance of almost all the “conventional” diagnostic tests against which Feno measurements were compared. This reflects the fact that, in unselected patients, most will have mild disease with normal lung function. In this setting Feno measurements may therefore be more relevant than traditional lung function tests. Interestingly, the combination of a raised Feno (>33 ppb) and abnormal spirometry (FEV1 <80% predicted) provides even greater sensitivity (94%) and specificity (93%) for the diagnosis of asthma.46,47

It is important to remember that patients may fulfil conventional clinical criteria for the diagnosis of asthma and yet Feno levels will be normal, especially in non-atopic subjects. Normal values do not exclude the diagnosis of asthma. Measuring AHR may reveal a positive clinically relevant result. Thus, Feno measurements complement AHR rather being a substitute for it, both in population surveys48 and in patients with asthma.49 This highlights further the fact that the asthma phenotype is heterogeneous, and that Feno measurements provide a perspective on only one aspect of the “asthma syndrome”.

Non-specific respiratory symptoms

Feno measurements have a wider role in assessing patients with undiagnosed chronic respiratory symptoms. There is a broad differential diagnosis in such patients depending on age. It includes eosinophilic bronchitis, cough variant asthma, post-viral bronchial hyperresponsiveness, postnasal drip and other ENT problems, gastro-oesophageal reflux disease, vocal cord dysfunction, primary hyperventilation syndrome, and COPD. In children, recurrent wheezy bronchitis, cystic fibrosis, congenital abnormalities of the airways or lungs, and primary ciliary dyskinesia also need to be considered.

FEno measurements may also permit the clinician to anticipate treatment responses. For eosinophilic bronchitis and cough variant asthma, which are characterised by eosinophilic airway inflammation and increased Feno levels, a positive response to a trial of steroid treatment is likely.11,50 On the other hand, for other diagnoses—for example, vocal cord dysfunction presenting as “asthma” which clinicians often treat empirically with steroids with little meaningful benefit51—it is just as helpful to have a low normal Feno level indicating a condition which is not characterised by eosinophilic airway inflammation and, in turn, is less likely to respond to steroids.

Preschool children

Given that spirometry and sputum induction cannot easily be performed in preschool children, a non-invasive measurement of airway inflammation is potentially very useful. As the single breath technique for measuring Feno in this age group is not suitable in preschool children, several alternatives have been developed, varying from modifications of the standard online technique to offline tidal breathing methods without flow control.52–61 In general, these techniques (which were reviewed by an ERS/ATS Task Force) are less sensitive in discriminating between asthmatic and non-asthmatic subjects.62,63

Evidence as to the overall diagnostic usefulness of Feno measurements in young children is mixed. In an unselected population of preschool children too young to perform spirometric tests, Feno performed poorly in distinguishing between asthma and non-asthma. Differences in Feno values between atopic children, children with doctor diagnosed asthma, and healthy children were less pronounced than in older subjects.64 However, when used in selected children, the performance characteristics are somewhat improved.65

In the differential diagnosis of non-specific respiratory symptoms, the same issues are encountered. Baraldi et al studied a group of 13 young children with recurrent wheeze and compared their Feno values with those of nine healthy controls and six children with a first episode of wheezing.53 Exhaled air was collected offline in a bag during tidal breathing without flow control. During an acute episode, Feno was significantly higher in those with recurrent wheeze than in controls, while in children with their first episode of wheezing Feno levels did not differ from normal children. These data are in keeping with those of Ratjen et al60 who measured peak Feno values online in mixed exhaled air (from mouth and nose).

A test that might allow better targeting of anti-inflammatory treatment, particularly in preschool children, would be very helpful. Feno is a promising tool in this regard. Treating infants and young children with recurrent wheeze and increased Feno levels with corticosteroids reduced Feno to normal or near normal values.53,66 Also, montelukast reduced Feno values in young children with early onset asthma.67,68

Influence of atopy

Several epidemiological studies have confirmed that Feno levels are raised in atopic subjects, whether or not they have significant lower respiratory tract symptoms.18,20,64,69–72 There is also a strong correlation between Feno levels and total as well as antigen specific IgE.71,72 This overall picture may be explained by the fact that even asymptomatic atopic patients may have mild airway inflammation,13,73 giving rise to increased Feno levels.

It has been suggested that the usefulness of Feno measurements may be limited only to atopic subjects, but we disagree with this view. Firstly, not all atopic individuals are identified using skin prick testing—that is, the label “non-atopic” may be falsely negative. Secondly, the presence of a low/normal Feno level in patients with chronic respiratory symptoms may be equally helpful in pointing the clinician away from the diagnosis of an atopic condition. In practice, when raised Feno levels are encountered in atopic subjects, additional investigations or treatment should be based on a history of significant symptoms. There is little evidence at present to support intervention in asymptomatic individuals.

Chronic obstructive pulmonary disease (COPD)

FEno levels are inconsistent in patients with COPD. This may be due to the confounding effect of current smoking or it may reflect the heterogeneity of underlying airway inflammation. Some studies report no significant change in Feno levels compared with controls,74,75 while others report that levels are increased.76 More recent evidence suggests that measuring alveolar rather than airway Feno may yield more important information,77 but at present this is technically demanding and beyond the scope of routine laboratory testing.

In older patients (>45 years) with fixed airflow obstruction, physiological tests alone are unhelpful in distinguishing those with asthma who would otherwise be classified as having COPD. Fabbri et al17 have shown that patients with historical evidence of asthma have eosinophilic airway inflammation in association with raised Feno levels. Earlier, Papi et al78 reported that increased sputum eosinophils and Feno levels occur in COPD patients with greater degrees of bronchodilator reversibility.

Perhaps the most important question is not whether the diagnostic label is accurate, but whether the response to anti-inflammatory treatment can be predicted. The data provided by Brightling et al,26 in which induced sputum eosinophil counts were used as the predictor, are encouraging. In that study 22 of 67 patients with COPD whose induced sputum eosinophil count was in the uppermost tertile (>4.5%) had significant symptomatic as well as physiological improvements with oral prednisone. Using Feno measurements, Zietkowski et al76 showed that the increase in post-bronchodilator FEV1 after 2 months of open label treatment with inhaled budesonide 800 μg/day was strongly correlated (r = 0.73, p = 0.0003) with baseline Feno levels in 19 ex-smoking patients with COPD. However, statistically significant correlations are not the same as predictive accuracy. There is a need for more work to be done to establish the exact role of Feno measurements in assessing COPD. This will require larger randomised controlled studies.

OTHER DISEASES IN WHICH FEno MAY HAVE A ROLE

Besides the common airways diseases, Feno measurements may have a role in the assessment of several other respiratory and non-respiratory conditions (table 1).

Table 1

 Respiratory and non-respiratory conditions in which Feno measurements may have a role in diagnosis

Cystic fibrosis (CF)

In patients with CF, Feno measurements have not been found to be clinically helpful. Values are usually normal or low.108–110 There are several possible explanations. Firstly, there is decreased expression of nitric oxide synthase (NOS-2) in patients with CF.116,117 Secondly, increased levels of nitrite are found in the breath condensate of patients with CF, suggesting trapping and metabolism of NO in secretions and mucus in CF airways.118,119

Primary ciliary dyskinesia (PCD)

Feno levels are significantly lower in patients with PCD than in healthy individuals, although with some overlap.111,120,121 Moreover, nasal NO (nNO) is extremely low in patients with PCD of all ages, and discriminates fully between affected and unaffected individuals. Measurement of nNO is likely to become the screening tool of choice.122 The diagnostic sensitivities and specificities of nNO for PCD range from 89% to 100% and from 97% to 100%, respectively. Low Feno and nNO levels may also be found in subjects with non-PCD bronchiectasis and sinus disease.112,121

Again, there are several possible explanations. Firstly, there may be decreased NOS activity. Administration of l-arginine as a substrate for NO increases nasal and exhaled NO formation in PCD, although not to normal values.123,124 This favours decreased NOS activity as a mechanism. Secondly, mucus may impair the diffusion of NO from the sinuses to the nasal cavity or from epithelial cells to the airway lumen, or may alter NO elimination.125 However, even in young infants with PCD, nNO is low, favouring the first explanation.126

NO is probably involved in stimulating ciliary motility.127 Nasal NO may also play a role in non-specific host defences, including direct toxic effects on micro-organisms.128 Reduced endogenous NO production and damage to NO producing cells may therefore contribute to recurrent airway infections.

Lung transplantation

Feno levels are increased in post-transplant patients with unstable lung function.87,88 More recent studies have investigated whether sequential Feno measurements can identify patients with progressive bronchiolitis obliterans syndrome (BOS).89,90 In a study by Brugiere et al,90 mean feno levels were twice as high in patients with progressive BOS than in those with or without BOS whose lung function remained stable over 14 months. Verleden et al89 evaluated the performance characteristics of Feno measurements over 2 years and obtained sensitivity, specificity, positive and negative predictive values for BOS of 92%, 84%, 80%, and 94%, respectively. The cut point used was 15 ppb (at an expiratory flow rate of 200 ml/s). This was equivalent to the upper limit of the 95% confidence interval for mean Feno levels in stable transplant patients. Interestingly, increased Feno levels preceded the changes in lung function by approximately 9 months. Although promising, the exact role of Feno measurements in post-transplant monitoring is not yet established.

FEno MEASUREMENTS IN THE MANAGEMENT OF CHRONIC ASTHMA

Two important questions have emerged regarding Feno measurements in the ongoing management of asthma:

  • Does Feno have prognostic significance?

  • Can Feno be used to guide treatment decisions relating to anti-inflammatory treatment?

Predicting exacerbations

Asthma is characterised by relapses and remissions, with deterioration in control provoked by a number of triggers as well as due to poor compliance with anti-inflammatory therapy. There is a perceived need for an objective measurement which might provide warning of impending deterioration or the need to change treatment. Peak flow measurements have been used to fulfil this role, but with limited success because changes in peak flow largely coincide with deteriorating symptoms rather than precede them.

Overall, the prognostic value of Feno measurements to predict deteriorating asthma appears limited. In a small study involving a steroid reduction protocol, Jatakanon et al129 reported that changes in sputum eosinophils were superior to Feno measurements in predicting loss of control. In a study by Jones et al37 measurements of AHR to hypertonic saline, sputum eosinophils, and Feno measurements all ranked similarly as predictors of control in 78 asthma patients following ICS withdrawal. Sensitivities ranged from 21% (eosinophils >4%) to 65% (Feno >10 ppb at a flow rate of 250 ml/s), although positive predictive values were in the range of 80–88%. Interestingly, the measurement of changes in these measurements was only marginally better than using single measurements. An increase in Feno of 60% was deemed to be optimum, but this was only 50% sensitive with a positive predictive value of 83%. These studies used a steroid withdrawal protocol to mimic a clinical exacerbation and are not necessarily ideal. In a much smaller study Harkin et al130 reported that, in routine practice, increased levels of Feno predicted an exacerbation within the following 2 weeks. It may be that, with the advent of portable monitoring, daily Feno measurements may prove to be beneficial in anticipating deteriorating asthma. However, as yet no data are available.

Predicting the outcome of ICS withdrawal in stable asthma

A relevant question is whether markers of airway inflammation can be used to predict the successful reduction or withdrawal of ICS treatment. In studies by Leuppi et al131 and Deykin et al,132 while sputum eosinophil counts (>0.8%) were highly predictive of subsequent loss of asthma control over periods of 6 months and 16 weeks respectively, no prognostic significance could be derived from Feno measurements. In the first study,131 baseline rather than sequential Feno values were used in the calculations. In the second, the number of patients in whom Feno values were obtained was limited, making valid comparisons difficult.132 In the study by Zacharasiewicz et al133 the negative predictive value of sputum eosinophils (at a cut point of 0%) was 100%—that is, treatment reduction/withdrawal was 100% successful (during the subsequent 8 weeks) when sputum eosinophilia was absent. A negative predictive value of 92% was obtained for Feno at a cut point of 22 ppb or less. Focusing on Feno, Pijnenburg et al134 reported that, following steroid withdrawal in currently asymptomatic children, Feno levels 2 and 4 weeks later were highly predictive of relapse during the subsequent 24 weeks of follow up, with a cut point for Feno of 49 ppb providing best predictive accuracy—that is, Feno levels above this threshold predicted likely asthma relapse.

Taken together, we can conclude that sputum eosinophil counts (>1%) probably offer superior prognostic accuracy when evaluating whether or not patients require ongoing ICS treatment. Furthermore, in circumstances where induced sputum cannot be obtained (in the majority of centres and in young children), a high Feno level (>50 ppb) is likely to predict asthma relapse and a low Feno level (<20 ppb in children, <25 ppb in adults) is likely to predict asthma stability if measured at least 4 weeks after ICS treatment is reduced/withdrawn in a currently asymptomatic patient. The outcome in those with an intermediate result (Feno 20–50 ppb) is less certain.

Adjustment of ICS dose

Several studies have recently explored whether “inflammometry” can be used to optimise the dose of ICS treatment. Using induced sputum eosinophil counts, Green et al24 showed that a management strategy which prescribed a stepwise reduction in ICS dose if sputum eosinophils were <1% (or an increase in dose if >3%) reduced asthma exacerbations to 32% of the rate observed in the control group. In another randomised study Jayaram et al133 have shown that, when ICS treatment is adjusted to maintain sputum eosinophils below 2%, the risk of eosinophilic exacerbations was reduced significantly (by 49%), with the number requiring intervention with prednisone reduced by two thirds.135 Interestingly, in that study the benefits of the “inflammometry” strategy occurred predominantly in patients with moderate or severe asthma.

The underlying rationale for each of these studies is both plausible and desirable—that is, anti-inflammatory treatment should be adjusted to ensure minimum airway inflammation. Currently, clinicians respond to uncontrolled symptoms or impaired lung function assuming that this results from uncontrolled airway inflammation. But the correlation between airway inflammation and either symptoms13,136 or lung function is weak,15,137 so the use of these end points to guide treatment can only be regarded as second best. It is far more rational that the ICS dose should be adjusted to take account of the principal action of ICS treatment—namely, control of airway inflammation.

Two randomised controlled trials in which Feno measurements have been used to guide long term treatment with ICS have recently been published.32,138 In both studies significant but differing benefits were obtained using an Feno based algorithm. In a study of 94 adult asthmatic patients by Smith et al,138 a 40% reduction in ICS dose requirements was achieved using Feno measurements at a single cut point of 35 ppb without any significant difference in the rate of asthma exacerbations (fig 2). In a study by Pijnenburg et al32 comprising 85 children with atopic asthma, the cut point for Feno levels was similar at 30 ppb; there was no difference in cumulative ICS use between the Feno and control groups. However, in the Feno group there was a significant reduction in the severity of AHR, with a concomitant (but, for reasons of study size, non-significant) reduction in exacerbations requiring oral prednisone.

Figure 2

 Profile of dose distribution for mean inhaled fluticasone requirements over 12 months in 48 patients (conventional group) whose ICS dose was adjusted using a priori guidelines and 46 patients in whom the ICS dose was adjusted on the basis of Feno measurements (Feno group; cut point equivalent to 35 ppb). There was a highly significant difference between the two groups (p = 0.008).138

Variation in the dose titration protocols and the study end points may account for the apparent differences in the results. However, overall, these studies provide encouraging evidence regarding the use of Feno in this clinical setting, although for the time being they must be regarded as indicative rather than conclusive. Firstly, in the study by Smith et al138 a single cut point for Feno was used to prompt either an increase or a decrease in ICS dose. However, a “one size fits all” approach may not be appropriate in regular clinical practice. Furthermore, although single cut points are appropriate in “back titration” studies with a high starting dose of ICS, two cut points defining three management choices (increase, no change or decrease in dose) may be more effective. This concept was included in the algorithm used by Pijnenburg et al32 but, clearly, this is an area which requires further investigation. Secondly, although the cut point for Feno was similar in both studies, substantially different criteria were used to guide ICS dose adjustment in the control groups. Irrespective of where the cut points are set, these will significantly determine the outcome in any dose adjustment strategy. This may explain why differing yet beneficial outcomes have been achieved in the “strategy” groups in the studies reported to date.24,32,138

An important and as yet unresolved issue is whether Feno measurements should be used for both upwards as well as downwards ICS dose titration. Clearly, the withdrawal of unnecessary ICS treatment or reducing excessive doses is an important goal of Feno monitoring. This was the primary objective in the study by Pijnenburg et al.32 In the study by Smith et al138 ICS dose reduction was a secondary end point but proved to be the most important outcome. Thus, data to date suggest that dose reduction is a fairly achievable objective. However, in patients with persistently high Feno levels (>50 ppb), it remains to be determined whether increasing the dose of ICS further will prove to be successful. It is doubtful whether it is justified if the patient is asymptomatic. However, although abolishing symptoms is a valid objective, it is not the only one. Higher doses of ICS reduce the frequency and impact of asthma exacerbations, and it may be that persistently raised Feno levels represent the signal to prescribe higher ICS doses with this objective in mind. This is controversial. On the one hand, the effect of ICS dose increments in the higher range on Feno is often small. Also, it is often not possible to normalise Feno values in patients with persistently high Feno levels, even on maximum doses.139 Equally, the cost-benefit ratio for very high doses of ICS in asthma increases. Thus, until further studies have been completed, it seems prudent to put the emphasis on dose reduction when Feno levels are low (<25 ppb). Only if asthma is poorly controlled and issues of poor compliance and/or poor inhaler technique have been addressed should high Feno levels prompt an increase in ICS dose.

MEASURING AND INTERPRETING FEno

Obtaining Feno measurements using commercially available analysers is, like riding a bicycle, fairly straightforward when you know how. The majority are “online”—that is, the analysis of exhaled gas is immediate and a result is available within a few minutes. Updated recommendations have recently been published and the reader is strongly advised to refer to them for fuller information.62,63 Establishing reference values for Feno measurements has been an evolving issue and remains problematic. Most of the studies conducted up to the year 2000 did not include a standardised method for Feno measurement. This severely limits their generalisablity. Further, our knowledge of factors which affect Feno levels in health and disease has expanded steadily. Demographic factors include age (adult versus paediatric) and smoking status.140–142 There is still some disagreement about whether atopy143 and sex144–146 are consistently important.

Normal values and clinically important changes

Against this background, current guidelines do not yet specify “normal” values.63 However, several recent studies have attempted to provide reference ranges for adults142,144,145 and children.145 In the study by Olin et al146 the interquartile range for Feno in healthy adults was 11.9–22.4 ppb. In a study comprising 30 healthy non-atopic adult subjects, the upper limit of normal (mean plus two standard deviations) was 33.1 ppb.146 In the study by Buchvald et al145 in a population of 405 children, the upper limit of the 95% confidence interval was age dependent, ranging from 15.7 ppb at the age of 4 years to 25.2 ppb for adolescents. The reason for this age dependency is unknown, but may be related to increasing airway surface area with age, age dependent induction of NOS secondary to recurrent immunological stimulation, or the progressive reduction over time of a constant exhalation flow rate which is relatively high in younger children. For the time being at least, it is reasonable that upper limits of normal for healthy adults and school age children should be set at 33 ppb and 25 ppb, respectively.

The reproducibility of Feno measurements is excellent with a very high intraclass correlation (>0.9) for repeated within sitting measurements.146 This translates into a within subject standard deviation for repeated measurements of 2.1 ppb for adults146 and 1.6 ppb for children145 (both at a flow rate of 50 ml/s). Similar high degrees of reproducibility have been reported by other authors.86,147–149

In clinical practice, as far as repeated measurements are concerned, the population of greatest interest are patients with chronic asthma. In this group the coefficient of variation for within subject between sitting measurements ranges from 10.5%37 to 26%.148 There is no “normal range” for patients with asthma. However, anecdotally, it would appear that even when asthma is well controlled, non-asthmatic “normal” Feno levels are rarely achieved. Thus, it may be that Feno levels measured when asthma is stable become the baseline reference point for individual patients against which subsequent measurements are weighed.

Kharitonov et al146 have reported that an absolute change of >4 ppb is significant. Based on the study by Ekroos et al,148 a percentage change of >26% would be deemed statistically significant. But are absolute or percentage changes of >4 ppb or >26% clinically important? And, if not, what constitutes a clinically meaningful change? In the study by Jones et al37 the median change in Feno which occurred between stability and “loss of control” after withdrawal of ICS treatment was 16.9 ppb (mean 24.9), but with a very large range (−10 to +141 ppb). The predictive accuracy of a change from baseline of 60% or greater was limited. These data suggest that group mean data may not be helpful in determining a clinically relevant change in individual patients. Further work is required to address this issue, particularly as to the changes (absolute and percentage) which might be anticipated in patients experiencing an asthma exacerbation while continuing to take regular controller therapy.

Interpretation

Based on currently available data, we have recently developed two algorithms for interpreting feno results in day to day practice—one for diagnostic use and the other for ongoing asthma management. These are shown in tables 2 and 3.

Table 2

 Feno levels as an aid to diagnosis of chronic respiratory symptoms

Table 3

 Feno levels as an aid in the management of asthma

Diagnostic use is fairly straightforward (table 2). For regular monitoring in subjects with chronic asthma (table 3), raised Feno levels in a symptomatic patient indicate uncontrolled eosinophilic airway inflammation. This is most frequently due to poor compliance with anti-inflammatory treatment or poor inhaler technique rather than inadequate ICS dosing. Although poor inhaler technique resulting in inadequate drug deposition is a plausible reason for raised Feno values, in a study of asthmatic children on a median dose of 800 μg budesonide this could not be proven.139 Where Feno levels remain high despite a seemingly adequate inhaled drug regime, it is theoretically possible that overexpression of constitutive steroid resistant NOS may be the explanation. Alternatively, alveolar NO rather than airway NO may be the source150,151 and, in such circumstances, a better clinical response may be achieved using oral rather than inhaled treatment.152 This remains controversial. Only rarely does a persistently high Feno level indicate true steroid resistance.

A low Feno level implies the absence of eosinophilic airway inflammation and, assuming that the result is not confounded by current tobacco smoking which may reduce Feno levels by up to 60%, an alternative or additional diagnosis to atopic asthma should be considered if the patient is symptomatic. The more common examples include non-atopic (possibly neutrophilic) asthma, gastro-oesophageal reflux disease, rhinosinusitis with postnasal drip, and left ventricular dysfunction.

The information contained in tables 2 and 3 is intended for guidance only. Future studies may indicate the need for modifications. Also, as indicated previously, patients with different clinical phenotypes may have different baseline values and different “target” Feno levels may be appropriate. This is because even when asthma is stable, Feno levels may remain high. As a strategy, evidence that “normalising” the Feno results in clinical benefit has not yet been documented. Rather, individualised “Feno typing” and cut points may be required. In some steroid dependent patients with asthma we have found it appropriate to devise a “sliding scale” which relates oral prednisone dose requirements to changes in Feno.

CONCLUSIONS

Feno measurements offer a step forward in the assessment of airways disease. As an “inflammometer”, Feno provides the clinician with hitherto unavailable information regarding the nature of underlying airway inflammation, thus complementing conventional physiological testing, including the measurement of AHR. Feno measurements are easy to perform, reproducible, and technically less demanding than induced sputum analysis. They are unreliable in current smokers and, when used diagnostically, in patients who have been taking inhaled or oral steroids recently.

Feno results require careful reference to the clinical context. In symptomatic patients, high Feno levels (>50 ppb) indicate significant airway eosinophilia which is likely to respond to ICS treatment. This appears to be independent of the diagnostic label. Further work is required to confirm how Feno measurements should be interpreted in patients with probable COPD. Present data provide support for the diagnostic use of Feno measurements in children with asthma-like symptoms, but in the very young more evidence is required. Whether or not Feno may be used to predict steroid response or guide ICS dose requirements in young children with recurrent wheeze is still unclear.

In patients with chronic and/or severe asthma, Feno levels are helpful to determine whether or not eosinophilic airway inflammation is currently active. Both high (>50 ppb) and low (<25 ppb) Feno levels may be used to predict outcomes in patients with a definite history of asthma currently in remission, and in whom withdrawal of ICS therapy is being undertaken. Again, depending on the level of symptoms, both high and low Feno levels offer the clinician information which may help to guide ICS dose adjustment decisions. As yet, however, much more work needs to be done before intermediate values based on group mean data can be used with complete confidence in this setting. The advantages of sequential individual data needs further study.