Airway resistance measured by the interrupter technique: normative data for 2–10 year olds of three ethnicities
  1. S A McKenzie,
  2. E Chan,
  3. I Dundas,
  4. P D Bridge,
  5. C S Pao,
  6. M Mylonopoulou,
  7. M J R Healy
  1. Royal London Hospital, UK
  1. Correspondence to:
    Dr S A McKenzie, 2nd floor, Fielden House, Royal London Hospital, Whitechapel, London E1 1BB, UK;
    s.a.mckenzie{at}qmul.ac.uk

Abstract

Background and Aims: The measurement of airway resistance using the interrupter technique (Rint) is feasible in preschool children and other subjects unable to undertake spirometry. This makes it potentially useful for the measurement of lung function in these groups. Commercial devices use different algorithms to measure pressure and flow from which Rint is derived. This study provides normative values for British children using devices from a single manufacturer.

Methods: Rint was measured in 236 healthy children of three ethnic groups (Afro-Caribbean and black African, Bangladeshi, and white British) aged 2–10 years using Micro Medical devices. Software in the devices calculated Rint from pressure measured by the two point, back extrapolation method from the pressure transient during valve closure, with flow measured just before valve closure.

Results: Rint is related to both age and height, but when age is allowed for there is not a significant relation with height. Neither gender nor any of the ethnicities studied was significantly related to Rint.

Discussion: These measurements in healthy children using this technique may be used as reference data for similar populations.

  • interrupter technique
  • normative data

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

The measurement of airway resistance by the interrupter method (Rint) has been developed by a number of groups for clinical use and as a research tool. The method is particularly suitable for subjects who cannot undertake standard lung function tests, such as preschool children and schoolchildren who are unable to undertake spirometry.1–7 The ability to monitor the progress of lung disease and to measure the effect of therapeutic interventions in these groups is quite limited. Much is already known about the technical aspects of the measurement of Rint, and its repeatability in young children.8

Commercial equipment for measuring Rint is now available. Undertaking the test is not complicated, even in a busy environment,6 and so it is possible to measure Rint outside specialist centres. It is therefore important to provide reference values for normal children using these devices. In Europe, manufacturers of interrupter equipment include Micro Medical (Micro Medical Ltd, Gillingham, UK) and Jaeger (E Jaeger, Wurzburg, Germany). Reference data for healthy children aged 6–11 years, used in the Micro Medical equipment, have been obtained from a group of Korean children.9 There are also data for Dutch children which include only a small number under 4 years,10 and for Italian preschool children.11 Reference data using the Jaeger equipment have been published from Danish children aged 2–7 years,12 and a Dutch population with the youngest subjects age 9 years.13 The two devices use quite different algorithms for the calculation of pressure and flow from which, in turn, Rint is calculated and so reference data for each are needed. Data from a small group of French children made with a non-commercial device have also been published.5

The purpose of this paper is to report measurements of Rint, using Micro Medical interrupter devices, in healthy east London children aged 2–10 years. This population is of particular interest as within it there are three large indigenous ethnic groups whose measurements can be compared. Our data will be compared to other published data.

SUBJECTS AND METHODS

Subjects

Subjects were children living in the east London districts of Tower Hamlets and City and Hackney. They were considered to have no respiratory problems if they met acceptable standards14: absence of acute disease or reported respiratory symptoms (such as cough or difficulty in breathing); chronic respiratory disease or symptoms; disorders, such as neurological disease, likely to affect the respiratory system; and no history of upper respiratory tract infection in the three weeks before the test. Ethnicity was classified as Afro-Caribbean (this group included black Africans), Bangladeshi, and white British. In the 1999 local census, these groups were represented in the study population as follows: Afro-Caribbean 10%, Bangladeshi 45%, white British 37%. Children of mixed ethnic parentage or from other ethnicities were excluded. Measurements of Rint were undertaken either on hospital premises or in schools. Those studied in hospital were patients who were attending for the ambulatory treatment of unrelated conditions such as fractures, or patients’ healthy siblings.

Methods

Rint was measured using two commercial devices (Microlab 4000 and MicroRint, Micro Medical Ltd, Gillingham, UK). In both devices flow is measured immediately before valve closure. Pressure is measured in both using a two point linear back extrapolation method. The time at valve closure (T0) is arbitrarily defined as the time at which 25% of the difference between the baseline pressure and the first clearly defined pressure peak preceding oscillation is reached.15 In both devices change in pressure is measured at the mouth following valve closure (Pmo(t)) by back extrapolating through 10 ms portions centred on T0+30 ms and T0+70 ms. The MicroRint extrapolates pressure to T0 and the Microlab 4000 to T0 +15 ms. Although the extrapolation to T0 + 15 ms would be expected to result in a higher pressure measurement and thus higher Rint measurements, it has been shown that measurements in children using the two devices show no systematic bias.16 This probably reflects a coefficient of variation of each measurement of about 15%,17 which is likely to obscure small differences in Rint. As part of our own laboratory’s quality control procedures, measurements have been made with both devices in 60 children aged 2–10 years with a wide range of Rint. The mean ratio of measurements made with the two devices, Microlab 4000/MicroRint, is 0.99 (95% confidence interval 0.95 to 1.03), suggesting there is no important systematic bias. Measurements were made in expiration. We have shown that these are similar to those in inspiration.17

Measurements were made as described previously.6 Subjects were seated in a similar, comfortable position. They breathed quietly through a cardboard mouthpiece (2.7 cm diameter or, for some of the younger children, 2.0 cm diameter) with the nose clipped, the cheeks and pharynx supported by the technician, and the neck slightly extended. After a period of quiet breathing, in response to a trigger during expiration at peak tidal flow, a single shutter closed automatically within 10 ms for 100 ms duration (manufacturer’s specification). Values were considered acceptable when the Pmo(t) was of consistent shape.3 At least six acceptable values of Rint were obtained and the mean of these values was considered the measurement. Subjects were unable to anticipate the trigger but could hear the shutter closing. Attempts were not accepted if breathing was irregular or the child was restless. All new operators were trained in the method so that interrater reliability was acceptable before measurements were made in subjects.6

Height was measured in stockinged feet using calibrated wall fixed stadiometers (Castlemead, Welwyn Garden City, Hertfordshire, UK) and portable, temporarily fixed stadiometers (Raven Equipment Ltd, Dunmow, Essex, UK).

Consent was obtained with an interpreter when necessary. The study was approved by the local ethics committee.

Data analysis

Using the UK cross sectional reference data,18 heights of subjects were expressed as z scores for age and compared by one way analysis of variance.

Measurements of Rint were transformed (log10) to produce a constant variance.19 The relations of log10Rint and height, age, gender, and the three ethnicities were described by simple and multiple linear regression. The widths of the confidence intervals describe the adequacy of the numbers of subjects studied to identify differences between genders and ethnicities.

RESULTS

There were 130 boys (median age 6.0 years, range 2.7–10 years) and 106 girls (median age 5.7 years, range 2.7–9.9 years). There were 39 Afro-Caribbean children (median age 5.4 years), 118 Bangladeshis (median age 6.1 years), and 79 white British (median age 5.0 years). Figure 1 shows the age distributions for subjects of each ethnicity. Seventy seven children had measurements made in schools, the remainder on hospital premises. Forty five measurements were made using the Microlab 4000 and 191 using the MicroRint.

Relation of height and ethnicity

Table 1 shows the mean z scores for height in the three ethnic groups. The differences are statistically significant (F2232 = 5.03, p = 0.007).

Table 1

Heights as z scores for age

Relation of Rint with age, height, gender, and ethnicity

Table 2 shows the results of multiple regression of log10Rint on height, age, gender, and ethnicity. It will be seen that the effects of gender and ethnicity were small and non-significant when age and height have been allowed for. Table 3 shows the results of regression on height and age. It appears that once age has been allowed for there is not a significant relation with height, but the reverse is not the case.

Table 2

Multiple regression of log10Rint on age, height, gender, and ethnicity

Table 3

Multiple regression of log10Rint on age and height

Table 4 gives the simple regression equations relating log10Rint to age and height. Measurements are plotted against age and height in figs 2 and 3. Graphical examination of the residuals showed that those for age were close to normality and had a constant variability across the range. There was no evidence of curvature. The regression on height showed significant curvature.

Table 4

Simple regressions of log10Rint on age and height

Figure 2

Measurements of Rint related to age.

Figure 3

Measurements of Rint related to height.

The mean ratio of Rint measurements girls to boys corrected for age, height, and ethnicity is 0.99 (95% CI 0.93 to 1.05). The mean ratio of Rint measurements in Bangladeshis, corrected for age, height, and gender, to those in white British is 1.00 (95% CI 0.93 to 1.07) and those in Afro-Caribbeans compared to white British is 1.02 (95% CI 0.93 to 1.12).

Comparison with published data

Published normative data measured with commercial devices are plotted in figs 4 and 5.9–13 The data from the Korean children were from measurements made using the Microlab 4000 using the same algorithm for calculating Rint as in this study. These provide the “predicted normal values” supplied with the Micro Medical interrupter equipment. Data for airway resistance measured by plethysmography are included for comparison.20

Figure 4

Published regression lines for age.

Figure 5

Published regression lines for height.

DISCUSSION

This is the first study which measures Rint in healthy children over the age range 2 to 10 years and examines the effect of ethnicity in addition to age, height and gender. We chose a population of predominantly prepubertal children as puberty may have an independent effect on Rint and would be too difficult to assess in children in a study such as this.

Our controls met standard criteria for children who are controls for respiratory disease.14 We did not ask about tobacco smoke exposure or smoking during pregnancy as it would have been impractical to validate this. Nor did we ask about family history of atopy or respiratory disease.

Height, age, gender, and ethnicity

We have presented simple regression equations, using both height and age, for comparison with other studies. We chose an exponential model because the scatter was homoscedastic after logarithmic transformation. Conventional lung function measurements such as FEV1 and FVC are usually standardised against height. In our data, height and age predicted Rint to the same extent. However, there was a small effect of age after height had been allowed for but the reverse was not the case. In addition, the age relation was linear whereas the height regression showed significant curvature even after the log transformation of Rint. Of the two, age will usually be the easier to ascertain, especially in field studies where accurate height measurement may be difficult. This is also true for measurements in physically disabled children. For practical purposes, we therefore recommend standardising Rint measurements against age, at least in prepubertal children.

No significant difference in Rint between boys and girls has been identified in other studies.9,12,13 In our study the 95% CI for the ratio of girls’ to boys’ measurements was 0.93 to 1.05, suggesting any difference is, at the very most, 6%. In infants born at term, respiratory resistance tended to be lower in girls than boys,21 but there is no information about measurements in the second year of life.

The effect of ethnicity was evaluated for the three main ethnic groups in east London. The Bangladeshi children studied here are all of Bangladeshi parents and are perceived to be a discrete ethnic group. The ethnicities of the black and white groups are likely to be less homogeneous. Nevertheless, using this classification the groups could be distinguished by height. We did not exclude children if their mothers smoked during pregnancy. European guidelines do not consider maternal smoking when reference values for lung function tests in children are compiled.22 About 40% of white British mothers in the area studied are believed to smoke during pregnancy23; fewer African and Afro-Caribbean mothers smoke, and the Bangladeshis rarely smoke. Smoking in pregnancy is well known to adversely affect lung function in infancy, probably reflecting smaller airways.23 Rint, however, was not significantly affected by these ethnicities.

Other studies

Reference values from different workers can reflect different techniques and different populations.22 The reference data used by Micro Medical in their equipment are from a large group of Korean children.9 Measurements were technically similar to ours, although only three values rather than six were used to calculate the measurement. Rint appears to be significantly lower in the Korean children. This could be the result of a genetic influence or simply reflect a healthier population.

The Danish equations12 have been derived using the “opening” interrupter algorithm.24 Using the Jaeger equipment, valve closure was at 50 ml of inspired volume,2 and pressure was measured over the last 5 ms of an 80 ms occlusion. Flow was measured over 5 ms, 70 ms after the valve opened. Examination of the Pmo(t) transient suggests that the pressure immediately before the valve opens will be higher than that calculated by two point back extrapolation.3 If flow is similar before and after occlusion, this would explain the higher measurements in the Danish study. In addition the Danish measurements were made with subjects breathing through a mask with an integral mouthpiece. Recent work has suggested that measurements made in this way are higher than those made with only the mouthpiece.25 Van Altena and Gimeno,13 using the Jaeger device in older children and adults, measured the pressure at the end of a 100 ms occlusion with valve closure at a flow of 0.6 l/s. Although their data for children suggest higher values than the other data sets, only 16 values for children under 11 years were included. The French measurements5 were made using a laboratory device. The valve closed at mid-tidal volume and pressure was measured by a two point, back extrapolation method at different points in the Pmo(t) transient from those used here. Flow was measured before occlusion. In this study, we have used commercially available devices. Most clinicians are likely to do the same.

Summary

We have measured Rint in a group of healthy British children aged 2–10 years and shown that measurements were not affected by gender or by the ethnicities within the group. As has been previously recommended,22,26 reference values from a particular population can be used for other populations if a representative sample from that population compares well. The absence of important ethnic differences in this study suggests that these data could be used for similar populations, using the same technique. Standardisation of the technique will further refine it so that interlaboratory comparisons of measurements can be made.27 The value of these data for clinical use will depend on how well Rint distinguishes healthy from sick children and the repeatability of the method. Measurements of bronchial hyperresponsiveness and bronchodilator responsiveness using the technique in preschool children have suggested that it may have a place.28,29 The repeatability of the measurement over time has been described in only a few subjects.10,11 Until more is known about this, no claims can be made of the value of Rint for following patients with chronic illness.