Table 3

 Predictive value of automated analysis of the Remmers sleep recorder

Diagnostic cut-offROC areaSensitivity (%)Specificity (%)Patients correctly classified (%)LR+LR−
RDISuz−, diagnostic cut-off value of the respiratory disturbance index (RDI) calculated by the Suzanne recorder based on respiratory events associated with a fall in Spo2 of at least 4%; RDISuz, diagnostic cut-off value of the RDI calculated by the Suzanne recorder; RDISuz+, diagnostic cut-off value of the extended RDI calculated by the Suzanne recorder and incorporating respiratory events associated with autonomic arousals; ROC area. area under the receiver operating characteristic curve; LR+, likelihood ratio for a positive test result; LR−, likelihood ratio for a negative test result.
RDISuz−
    ⩾50.9594.174.485.13.680.08
    ⩾100.9690.982.085.15.040.11
    ⩾150.9995.287.789.47.720.05
RDISuz
    ⩾50.8575.381.076.63.950.30
    ⩾100.9067.987.876.65.570.37
    ⩾150.9162.896.180.916.00.39
RDISuz+
    ⩾50.8473.783.375.54.420.31
    ⩾100.8260.696.471.317.00.41
    ⩾150.9451.997.571.320.70.49