Thorax 2007;62:565-568 doi:10.1136/thx.2006.072892
  • Lung cancer

Diagnosis of lung cancer by the analysis of exhaled breath with a colorimetric sensor array

  1. Peter J Mazzone,
  2. Jeffrey Hammel,
  3. Raed Dweik,
  4. Jie Na,
  5. Carmen Czich,
  6. Daniel Laskowski,
  7. Tarek Mekhail
  1. The Cleveland Clinic, Cleveland, Ohio, USA
  1. Correspondence to:
    Dr Peter J Mazzone
    Department of Pulmonary, Allergy, and Critical Care Medicine, The Cleveland Clinic, 9500 Euclid Avenue, A90 Cleveland, OH 44195, USA; mazzonp{at}ccf.org
  • Received 4 October 2006
  • Accepted 3 January 2007
  • Published Online First 27 February 2007

Abstract

Background: The pattern of volatile organic compounds (VOCs) in the exhaled breath of patients with lung cancer may be unique. New sensor systems that detect patterns of VOCs have been developed. One of these sensor systems, a colorimetric sensor array, has 36 spots composed of different chemically sensitive compounds impregnated on a disposable cartridge. The colours of these spots change based on the chemicals with which they come into contact. In this proof of principle study, the ability of this sensor system to detect a pattern of VOCs unique to lung cancer is assessed.

Methods: Individuals with lung cancer, those with other lung diseases and healthy controls performed tidal breathing of room air for 12 min while exhaling into a device designed to draw their breath across a colorimetric sensor array. The colour changes that occurred for each individual were converted into a numerical vector. The vectors were analysed statistically, using a random forests technique, to determine whether lung cancer could be predicted from the responses of the sensor.

Results: 143 individuals participated in the study: 49 with non-small cell lung cancer, 18 with chronic obstructive pulmonary disease 15 with idiopathic pulmonary fibrosis 20 with pulmonary arterial hypertension 20 with sarcoidosis and 21 controls. A prediction model was developed using observations from 70% of the subjects. This model was able to predict the presence of lung cancer in the remaining 30% of subjects with a sensitivity of 73.3% and a specificity of 72.4% (p = 0.01).

Conclusions: The unique chemical signature of the breath of patients with lung cancer can be detected with moderate accuracy by a colorimetric sensor array.

Footnotes

  • Competing interests: None.

  • Chemsensing (Champagne, Illinois, USA) provided the sensor arrays, breath collection and analysis equipment free of charge. They were not involved in the statistical analysis or writing of this manuscript.

This Article

  1. All Versions of this Article:
    1. thx.2006.072892v1
    2. 62/7/565 most recent

Services

  1. Request permissions

Social bookmarking

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of Thorax.
View free sample issue >>

Free archive
The full back archive is now available for Thorax. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.