Thorax 2002;57:258-262 doi:10.1136/thorax.57.3.258
  • Original articles

Effect of assist negative pressure ventilation by microprocessor based iron lung on breathing effort

  1. M Gorini,
  2. G Villella,
  3. R Ginanni,
  4. A Augustynen,
  5. D Tozzi,
  6. A Corrado
  1. Respiratory Intensive Care Unit, Careggi Hospital, Firenze, Italy
  1. Correspondence to:
    Dr M Gorini, Via Ragazzi del 99 60, 50141 Firenze, Italy;
    mgorini{at}qubisoft.it
  • Accepted 27 September 2001
  • Revised 10 August 2001

Abstract

Background: The lack of patient triggering capability during negative pressure ventilation (NPV) may contribute to poor patient synchrony and induction of upper airway collapse. This study was undertaken to evaluate the performance of a microprocessor based iron lung capable of thermistor triggering.

Methods: The effects of NPV with thermistor triggering were studied in four normal subjects and six patients with an acute exacerbation of chronic obstructive pulmonary disease (COPD) by measuring: (1) the time delay (TDtr) between the onset of inspiratory airflow and the start of assisted breathing; (2) the pressure-time product of the diaphragm (PTPdi); and (3) non-triggering inspiratory efforts (NonTrEf). In patients the effects of negative extrathoracic end expiratory pressure (NEEP) added to NPV were also evaluated.

Results: With increasing trigger sensitivity the mean (SE) TDtr ranged from 0.29 (0.02) s to 0.21 (0.01) s (mean difference 0.08 s, 95% CI 0.05 to 0.12) in normal subjects and from 0.30 (0.02) s to 0.21 (0.01) s (mean difference 0.09 s, 95% CI 0.06 to 0.12) in patients with COPD; NonTrEf ranged from 8.2 (1.8)% to 1.2 (0.1)% of the total breaths in normal subjects and from 11.8 (2.2)% to 2.5 (0.4)% in patients with COPD. Compared with spontaneous breathing, PTPdi decreased significantly with NPV both in normal subjects and in patients with COPD. NEEP added to NPV resulted in a significant decrease in dynamic intrinsic PEEP, diaphragm effort exerted in the pre-trigger phase, and NonTrEf.

Conclusions: Microprocessor based iron lung capable of thermistor triggering was able to perform assist NPV with acceptable TDtr, significant unloading of the diaphragm, and a low rate of NonTrEf. NEEP added to NPV improved the synchrony between the patient and the ventilator.

Footnotes

    Register for free content


    Free sample
    This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of Thorax.
    View free sample issue >>

    Free archive
    The full back archive is now available for Thorax. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
    Register to access the free archive >>

    Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.