Acute anaphylaxis following midline catheterisation in a patient with cystic fibrosis

  1. C S Haworth,
  2. R McL Niven,
  3. A J Moorcroft,
  4. A Phillips,
  5. M E Dodd,
  6. A K Webb
  1. Adult Cystic Fibrosis Unit, Wythenshawe Hospital, Southmoor Road, Manchester M23 9LT, UK
  1. Dr A K Webb.
  • Received 26 January 1998
  • Accepted 8 May 1998

Abstract

A 16 year old male with cystic fibrosis experienced an acute life threatening anaphylactic reaction following the insertion of an Ohmeda Hydrocath™ midline peripheral venous catheter. The catheter was immediately withdrawn and treatment with intravenous adrenaline, hydrocortisone, chlorpheniramine, and colloid over a 24 hour period resulted in a gradual resolution of symptoms.

Patients with cystic fibrosis often require frequent treatment with intravenous antibiotics and these are now commonly administered via midline peripheral venous catheters (longlines). Longlines have significant advantage over cannulas in terms of duration of patency. Unfortunately, however, life threatening complications can occur. We report three cases of anaphylaxis associated with Ohmeda Hydrocath™ peripheral venous catheters.

Case report

A 16 year old male with cystic fibrosis attended clinic to start a routine course of home intravenous antibiotic therapy. He was colonised with Pseudomonas aeruginosa when he was four years old and since the age of six and a half had been treated with intravenous antibiotics on a three monthly basis, usually with meropenem via a Vygon longline. Prior to treatment his forced expiratory volume in one second was 102% of predicted. He had a past medical history of portal hypertension and oesophageal varices.

An Ohmeda Hydrocath midline peripheral venous catheter was inserted for the two week course of intravenous antibiotics. Three months earlier the patient had used this type of longline for the first time without complication. No subcutaneous anaesthetic was used. Approximately one minute after insertion, before the line had been flushed, the patient complained of nausea and breathlessness. He then deteriorated rapidly, displaying the signs and symptoms typical of acute anaphylaxis. On examination he was distressed and had facial swelling. His blood pressure was unrecordable. Examination of the chest revealed new widespread expiratory wheeze. Acute anaphylaxis was diagnosed and the longline was immediately withdrawn. Intravenous access was established with a cannula (Venflon™) and chlorpheniramine, hydrocortisone, adrenaline, and colloid were administered. His condition stabilised over a two hour period but an adrenaline infusion was required overnight to support his blood pressure. The following day he was discharged home.

Two further patients attending our clinic have experienced anaphylactic reactions following the insertion of an Ohmeda Hydrocath midline peripheral venous catheter. The reactions were characterised by facial swelling, hypotension, tachycardia, and bronchospasm. In one patient the reaction developed after the longline was flushed with heparinised saline and may therefore represent heparin induced histamine release. In both cases prompt removal of the longline and treatment with hydrocortisone, adrenaline, and chlorpheniramine resulted in complete resolution of symptoms. In each case the longlines were inserted using the same technique, but by a different physician.

In all three patients skin tests to water extracts from the line, guide wire, and gloves failed to produce a reaction. IgE specific antibodies to latex and ethylene oxide (the sterilising agent) were also normal.

Discussion

Systemic reactions to elastomeric hydrogel midline catheters have been reported.1-3 To our knowledge this is the first published account of a life threatening reaction after the insertion of an Ohmeda Hydrocath polyurethane midline peripheral venous catheter.

Considering the large number of longlines used in the treatment of cystic fibrosis and other conditions, these events are fortunately rare. However, those responsible for inserting longlines should be aware of the possible complications and have access to emergency resuscitation facilities.

References

This Article

Services

  1. Request permissions

Social bookmarking

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of Thorax.
View free sample issue >>

Free archive
The full back archive is now available for Thorax. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.