Article Text

Download PDFPDF

Effectiveness of a heat and moisture exchanger in preventing hyperpnoea induced bronchoconstriction in subjects with asthma.
Free
  1. T R Gravelyn,
  2. M Capper,
  3. W L Eschenbacher
  1. Department of Internal Medicine, University of Michigan Medical Center, Ann Arbor 48109-0026.

    Abstract

    The effect of a heat and moisture exchanger, a device with hygroscopic material for conditioning inspired air, on hyperpnoea induced bronchoconstriction was studied in nine non-smoking volunteers with asthma, aged 19-32 years. Each had previously shown an increase of at least 100% in specific airways resistance (sRaw) to isocapnic hyperpnoea with dry air. On two separate days the subject performed isocapnic hyperpnoea with dry air at 60-70 l min-1 for five minutes. Before, immediately after, and five minutes after completion of a test sRaw measurements were made. Heat and moisture exchangers were placed in the breathing circuit on one of the two days. All subjects had an increase in sRaw of 100% or more without the heat and moisture exchangers (average increase 300%) but were protected from bronchoconstriction with the devices in place (average increase 7%) (p less than 0.005). The exchanger's resistance to airflow was less than 1 cm H2O for flow rates of 100 l min-1. A heat and moisture exchanger designed as a facemask or mouthpiece may allow a person with asthma to exercise without the need for prophylactic drugs.

    Statistics from Altmetric.com

    Request Permissions

    If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.