Thorax. Published Online First: 24 January 2007. doi:10.1136/thx.2006.061234
Papers |
Predicitve Value Of Automated Oxygen Saturation Analysis For The Diagnosis And Treatment Of Obstructive Sleep Apnea In A Home-Based Setting
1 Centre Hospitalier de l'Université de Montréal (CHUM), Canada
* To whom correspondence should be addressed. E-mail: bellemaf{at}colba.net.
Accepted 20 August 2006
Abstract
Background: A portable monitor for the automated analysis of episodic nocturnal oxygen saturation or SpO2 (the Remmers Sleep Recorder, RSR) was proposed for the diagnosis of obstructive sleep apnea-hypopnea (OSAH). The purpose of this study was to compare the diagnostic performance of automated analysis with the manual scoring of polygraphic data by a more comprehensive respiratory monitor (the Suzanne recorder) used simultaneously in their intended home environment.
Methods: The respiratory disturbance indexes of the 2 monitors were compared in 94 consecutive adult patients suspected of having OSAH and who were deemed eligible for home-based investigation.
Results: The RSR overestimated the number of respiratory events associated with
4% fall in SpO2 by 13% (p<.005), but underestimated the number of apneas and hypopneas defined on the basis of respiratory variables alone or their association with
4% fall in SpO2 or autonomic arousals, by 38% to 48% (p<.0001). In addition to these significant biases, the limits of agreement in all instances were wide, indicating a poor concurrence between the 2 monitors.
Conclusion: The automated analysis of SpO2 with the RSR cannot be substituted for the manual scoring of polygraphic data with the more comprehensive respiratory monitor in the diagnosis of OSAH in an ambulatory home-based setting.
Keywords: Obstructive sleep apnea, diagnosis, oximetry, polysomnography, sleepiness
Relevant Articles
- Airwaves
- Wisia Wedzicha
Thorax 2007 62: 373a.[Extract] [Full Text] [PDF]
- Abbreviated or not abbreviated? Is it the right question?
- Frederic Sériès
Thorax 2007 62: 379-380.[Extract] [Full Text] [PDF]
This article has been cited by other articles:
-
Ruiz-Lopez, F. J., Fernandez-Suarez, B., Guardiola-Martinez, J., Vergara-LaHuerta, I., Latour-Perez, J., Lorenzo-Cruz, M.
(2009). Quality Control of the Ambulatory Polygraphy Using Automatic Analysis. Chest
135: 194-200
[Abstract] [Full Text] -
McNicholas, W. T.
(2008). Diagnosis of Obstructive Sleep Apnea in Adults. Proc Am Thorac Soc
5: 154-160
[Abstract] [Full Text] -
Series, F.
(2007). Abbreviated or not abbreviated? Is it the right question?. Thorax
62: 379-380
[Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.