Thorax

HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS REGISTER
[Advanced]

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Web-only Tables
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this link to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Add article to my folders
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Miller, M R
Right arrow Articles by Pedersen, O F
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Miller, M R
Right arrow Articles by Pedersen, O F
Thorax 2003;58:411-416
© 2003 BMJ Publishing Group & British Thoracic Society


RESPIRATORY PHYSIOLOGY

Inadequate peak expiratory flow meter characteristics detected by a computerised explosive decompression device

M R Miller1, P R Atkins2, O F Pedersen3

1 Department of Medicine, University of Birmingham, Birmingham, UK
2 Department of Electronic and Electrical Engineering, University of Birmingham, Birmingham, UK
3 Department of Occupational and Environmental Medicine, University of Aarhus, Aarhus, Denmark

Correspondence to:
Correspondence to:
Dr M R Miller, University of Birmingham, Department of Medicine, Selly Oak Hospital, Birmingham B29 6JD, UK;
m.r.miller{at}bham.ac.uk


ABSTRACT
Background: Recent evidence suggests that the frequency response requirements for peak expiratory flow (PEF) meters are higher than was first thought and that the American Thoracic Society (ATS) waveforms to test PEF meters may not be adequate for the purpose.

Methods: The dynamic response of mini-Wright (MW), Vitalograph (V), TruZone (TZ), MultiSpiro (MS) and pneumotachograph (PT) flow meters was tested by delivering two differently shaped flow-time profiles from a computer controlled explosive decompression device fitted with a fast response solenoid valve. These profiles matched population 5th and 95th centiles for rise time from 10% to 90% of PEF and dwell time of flow above 90% PEF. Profiles were delivered five times with identical chamber pressure and solenoid aperture at PEF. Any difference in recorded PEF for the two profiles indicates a poor dynamic response.

Results: The absolute (% of mean) flow differences in l/min for the V, MW, and PT PEF meters were 25 (4.7), 20 (3.9), and 2 (0.3), respectively, at PEF {approx}500 l/min, and 25 (10.5), 20 (8.7) and 6 (3.0) at {approx}200 l/min. For TZ and MS meters at {approx}500 l/min the differences were 228 (36.1) and 257 (39.2), respectively, and at {approx}200 l/min they were 51 (23.9) and 1 (0.5). All the meters met ATS accuracy requirements when tested with their waveforms.

Conclusions: An improved method for testing the dynamic response of flow meters detects marked overshoot (underdamping) of TZ and MS responses not identified by the 26 ATS waveforms. This error could cause patient misclassification when using such meters with asthma guidelines.


Keywords: peak expiratory flow; peak flow meters; lung function




This article has been cited by other articles:


Home page
Eur Respir JHome page
M. R. Miller, J. Hankinson, V. Brusasco, F. Burgos, R. Casaburi, A. Coates, R. Crapo, P. Enright, C. P. M. van der Grinten, P. Gustafsson, et al.
Standardisation of spirometry
Eur. Respir. J., August 1, 2005; 26(2): 319 - 338.
[Abstract] [Full Text] [PDF]


Home page
ThoraxHome page
C K Connolly and V Brusasco
Peak flow meters still useful but require consistency rather than accuracy
Thorax, January 1, 2004; 59(1): 82 - 83.
[Full Text] [PDF]


Home page
ThoraxHome page
V Brusasco
Usefulness of peak expiratory flow measurements: is it just a matter of instrument accuracy?
Thorax, May 1, 2003; 58(5): 375 - 376.
[Full Text]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS REGISTER
Terms and conditions relating to subscriptions purchased online  ¦  Website terms and conditions  ¦  Privacy policy
Copyright © 2003 BMJ Publishing Group Ltd & British Thoracic Society